Blood pressure measurements in research: suitability of auscultatory, beat-to-beat, and ambulatory blood pressure measurements

  • Debra J Carlson
  • , Gudrun Dieberg
  • , James R McFarlane
  • , Neil A Smart

Research output: Contribution to journalArticlepeer-review

15 Citations (Scopus)

Abstract

Objective The objective of this study was to validate the accuracy of beat-to-beat measurements with those taken with an aneroid sphygmomanometer by auscultatory method. A secondary aim was to explore differences between auscultatory and beat-to-beat blood pressure (BP) with daytime ambulatory BP measurements. Participants and methods A total of 46 participants, comprising 21 males, aged 47± 13 years, height 171± 8.5 cm and weight 82 ± 16.8 kg attended the Exercise Physiology Laboratory at the University of New England (Armidale, New South Wales, Australia). During the visit, participants had their BP – systolic BP (SBP) and diastolic BP (DBP) – measured using auscultatory methods and a Finometer. An ambulatory BP monitor was fitted during the same visit and worn for a minimum of 12 h. Results Auscultatory measurements were slightly higher than beat-to-beat for both SBP and DBP. There was no difference between auscultatory and beat-to-beat SBP with a mean difference of 0.23 mmHg (P = 0.87). There were disparities between auscultatory and beat-to-beat DBP, with a mean difference of 4.82 mmHg (P < 0.01). Daytime ambulatory BP was higher than both auscultatory and beat-to-beat measurements for both SBP and DBP, with P less than 0.001 for all measures. Conclusion There was a high level of reliability in the beat-to-beat SBP with that seen by auscultatory; however, there were disparities in DBP measurements using the same devices, which raise concerns over the accuracy of beat-to-beat DBP. Ambulatory systolic and diastolic measures were higher than beat-to-beat and auscultatory; however, they may be more suitable for monitoring diurnal changes in BP, depending upon the research model.
Original languageEnglish
Pages (from-to)18-23
JournalBlood Pressure Monitoring
Volume24
Issue number1
DOIs
Publication statusPublished - 1 Feb 2019

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Fingerprint

Dive into the research topics of 'Blood pressure measurements in research: suitability of auscultatory, beat-to-beat, and ambulatory blood pressure measurements'. Together they form a unique fingerprint.

Cite this